Perclose Proglide® Single Device Deployment

Refer to Instructions for Use for additional information.

Watch Now >

Single Device Deployment

Pre-Deployment
  • Flush Marker Lumen
  • Insert device over the guide wire
  • Remove guide wire when Guide Wire Exit Port is at skin level
  • Advance device until pulsatile flow is observed from Marker Lumen
Suture Deployment

1Advance device and lift Lever (Open Foot). In the vein, blood may only fill the Marker Lumen. Lift Lever to deploy the Foot

2Depress Plunger (Deploy Needles). Retract device until the Foot has apposition against the vessel wall. Stabilize the device to approximately 45°, maintaining slight upward force and depress Plunger. Cessation of blood from the Marker Lumen may be noted

3Pull Back Plunger (Deploy Suture). Pull suture taught and cut suture using the QuickCut®

4Lower Lever (Close Foot). Relax device and return Lever to the original position

Suture Management
  • Retract device until Guide Wire Exit Port is at skin level
  • Remove the suture limbs
    • Blue limb = Rail
    • White limb = Non-Rail
  • Insert guide wire into Guide Wire Exit Port
  • Remove device while pulling tension on Rail (blue)
  • Load Snare Knot Pusher (SKP) on to Rail (blue) suture
  • Apply slow consistent tension to advance knot down to vessel
  • Assess hemostasis and remove guide wire while maintaining tension on Rail suture
  • Load Snare Knot Pusher (SKP) on to Rail (blue) suture
  • Lock knot by pulling on Non-Rail (white) suture
  • Use right thumb to pull back the Thumb Knob on the Suture Trimmer to load both suture limbs
  • Advance Suture Trimmer down to vessel
  • Use right index finger to pull back the Red Lever; cut suture
Assess Hemostasis

Challenge and Confirm the Close

  • Have the patient bend leg/lift head and cough
  • If further knot tightening is required, repeat Suture Management steps
Please wait...
Please wait...

 

Are you a healthcare professional in the U.S. or outside of the U.S.?

U.S.   Outside of U.S.

Close

Sign Up For a Hands-On Demo

Please fill out the form below. An Abbott Sales Rep will be in touch shortly.

Yes, I'd like to be informed about Abbott products and be invited to symposia, events and more.

All information collected will be used for sending updates only and will not be shared with third parties. For more details, see our Privacy Policy.

Close

Request More Information

Please fill out the form below. An Abbott Sales Rep will be in touch shortly.

Yes, I'd like to be informed about Abbott products and be invited to symposia, events and more.

All information collected will be used for sending updates only and will not be shared with third parties. For more details, see our Privacy Policy.